Product Specifications
LIGHT SOURCE Quartz-halogen lamp 12V/20W
WAVELENGTH RANGE Automatic by 8 positions filter-wheel
7 standard filters: 340nm, 405nm, 492nm, 510nm, 546nm, 578nm, 630nm
1 position free for optional filter
Half bandwidth< 8 nm
Stray light : < 1.0% at 340nm
PHOTOMETRIC RANGE 0.0000 to 3.0000 ABS
Resolution : 0.0001 ABS
Stability : 0.002 A/hour
FLOWCELL Stainless steel with quartz window
Measuring volume 32 µL
Optical path 10mm
Aspiration volume programmable : 200~2000µL
Carryover < 1.0%
ANALYSIS METHOD End point , with or without reagent blank ,
End point ,with sample bland, with or without reagent blank
Kinetic , with or without linearity check
Two-point kinetic /fixed time , with or without reagent blank
Absorbance
Turbidimetry
Linear and non-linear calibration
PARAMETER SETTINGS Method
Wavelength
Temperature
Reagent blank y/n
Sample blank y/n
Lag time
Measurement time
Reaction time
Absorbance limit
Aspiration volume
Standards
Linearity check
Unit for results
MEMORY : 112 test profiles , 100,000 sample results
TEMPERATURE CONTROL : By means of Peltier elements 25°C, 30°C, 37°C ±0.1°C
INPUT Touch screen
DISPLAY 7 inch Large LCD display
PRINT Built-in thermal recorder
INTERFACE RS232 serial port
Input/Output : USB port can be used to upgrade software version
OPERATING ENVIRONMENT Temperature : 15°C ~30°C
Humidity ; ≤70%
POWER REQUIREMENT AC 220V / 110V ±10%, 50/60Hz
DIMENSION 420mm(H) x 380mm(W) x 170mm(D)